Register   |   Login
 Field Engineer Calibration:

The Wintronics Field Service System manages all your engineers' calibration needs - from equipment scheduling and tracking to shipping and calibration - and provides quality reports and certificates of calibration. 

 

 Calibration brochure:
 IMPORTANT LINKS:

CT13 Series Infrared Thermometer

The Universal Infrared Radiation Pyrometer for non-Contact Temperature Measurement.

Download the complete CT13 Brochure.

 Features:
  • Rugged stainless steel housing, NEMA 4
  • Wide temperature range from 0°C to 900°C
  • Very fast response times > 30ms (programmable)
  • Fields of view as small as 1 mm
  • Focus laser marks field of view area
  • High accuracy (0.7%) and long term stability
  • Chopped Radiation Method for very high temp. stability
  • µP based signal operation
All parameters configurable via serial interface:
  • Temperature measuring range
  • Analog output 0/4-20mA; actual or Min/Max-value
  • Integrated valley/peak picker w/ decay function
  • Emissivity setting
  • Reflected environmental temperature

 

 General Specifications:
Temperature range 0°C ... 900°C
Temperature resolution (NETD) Depends on measuring temperature and response time, typical value 0.2°C (at 300 ms, 100 °C, e = 1)
Accuracy (uncertainty) ± 0.8°C plus 0.8% of the difference between target and sensor head temperature
Long term stability Better than 0.01% of the absolute measured temperature per month
Field of view diameter From Ø 1 mm, depends on lens
Field of view marking Laser pointer as accessory for non transparent lenses
Pilot laser, built-in: aims the center of the field of view in any distance
(More mechanical pointers are available)
Laser function Time out or permanent operation, while flashing or continuous marking
Spectral ranges 2 - 2.7 µm,  3.9 µm,  4.9 - 5.6 µm,  7.93 µm,  7.5 - 8.2  µm,  8 - 14 µm
Programmable functions via serial interface Emissivity, Environmental temperature, Analog output, Function of analog output, Response time, Temperature unit, Valley/ Peak-picker with decay function, Laser operation
Emissivity 0.100 to 1,000 in 0,001-steps
Response time from 30 ms to 10 s (0,03, 0,1, 0,3, 1, 3, 10 s); Option: 5 ms available
Temperature unit °C, °F or K

Analog output (Hardware)

 

Analog output (Functions)

linear 0-20mA or 4-20mA, Zoom function for temperature span > 50 °C.

 

Actual value, Maximum value or Minimum value

Serial interface RS232-Interface, bi directional, 9,600 to 57,600 bps, for programming and data transfer
Thermal switch Monitors the instrument temperature
Power requirements 10.5-32 VDC or 10.8 VAC to 26.4 VAC, 48-400 Hz
Approx. 2.5 W
Permissible operating temperature

-25°C ... 60°C

With protective and cooling housing WK15 up to 300 °C

Storage temperature -40°C ... 85°C
Protective class, Weight IP68 (IEC), NEMA 6; 1.3 kg
Housing Stainless steel

Wintronics Inc., exclusive US agent for Heitronics Infrared Radiation Pyrometers

 

 CT13

WINTRONICS, INC.  •   50 Division Ave.  •  Millington, NJ 07946  •   908-647-0144  •  800-247-7976  •  fax: 908-647-8379  •  win@wintron.com