CT09.10 Data Sheet
-30 to 500 °C or 0 to 900 °C
Depends on measured temperature and response time, typical value 0.2°C (at 300 ms, 100°C, ε = 1)
+/- 1.0 °C plus 0.6% of the difference between target and sensor head temperature as a function of housing temperature: 0.01% / °C for housing temp. other than 25°C
Better than 0.01% of the absolute measured value per month
From 1 mm diameter, depends on lens (see above)
From 50 ms to 10 s (0.05; 0.1; 0.3; 1; 3; 10 s)
Linear 0 - 20 mA, or 4 - 20 mA, scalable temperature span => 50 °C
RS232-interface, bi-directional, 9.6 to 57.6 kbps, for programming and data transfer
15 VDC to 32 VDC, or 10 VDC to 15 VDC
Approx. 1.6 W
-25 to +70 °C
With protective and cooling housing WK11 up to 300 °C
-40 to +85 °C
Stainless Steel,
Protective Class IP65 (IEC), (NEMA 4 equivalent)
Wintronics Inc., exclusive US agent for HEITRONICS Infrared Radiation Pyrometers
Wintronics Calibration provides the highest level of measurement service available today to ensure the exacting quality you demand from your measurement equipment.
NIST
HEITRONICS
PTB (National Metrology Institute of Germany)
Wintronics, Inc. is an ISO/IEC 17025:2017 Accredited Laboratory that performs Calibration traceable to International Standards.
Download ISO Certificate
Wintronics, Inc.
50 Division Avenue
Millington, NJ 07946
800-247-7976908-647-0144
fax: 908-647-8379
win@wintron.com
Contact Wintronics