CT13 Series Infrared Thermometer

 

 

 CT13

 


The Universal Infrared Radiation Pyrometer for Non-Contact Temperature Measurement.
 

 

CT13-Series Brochure

 

 

CT13-Series Datasheet

 

 

 

 

 

Features of the CT13 Infrared Radiation Thermometer

 

  • Rugged stainless steel housing, NEMA 6
  • Wide temperature range from 0°C to 900°C
  • Very fast response times > 30ms (programmable)
  • Fields of view as small as 1 mm
  • Focus laser marks field of view area
  • High accuracy (0.7%) and long term stability
  • Chopped Radiation Method for very high temp. stability
  • µP based signal operation

   
 

All parameters configurable via serial interface:

 

  • Temperature measuring range
  • Analog output 0/4-20mA; actual or Min/Max-value
  • Integrated valley/peak picker w/ decay function
  • Emissivity setting
  • Reflected environmental temperature

General Specifications

Temperature range 0°C ... 900°C
Temperature resolution (NETD) Depends on measuring temperature and response time, typical value 0.2°C (at 300 ms, 100 °C, e = 1)
Accuracy (uncertainty) ± 0.8°C plus 0.8% of the difference between target and sensor head temperature
Long term stability Better than 0.01% of the absolute measured temperature per month
Field of view diameter From Ø 1 mm, depending on lens
Field of view marking Laser pointer as accessory for non transparent lenses
Pilot laser, built-in: aims the center of the field of view in any distance
(More mechanical pointers are available)
Laser function Time out or permanent operation, while flashing or continuous marking
Spectral ranges 2 - 2.7 µm,  3.9 µm,  4.9 - 5.6 µm,  7.93 µm,  7.5 - 8.2  µm,  8 - 14 µm
Programmable functions via serial interface Emissivity, Environmental temperature, Analog output, Function of analog output, Response time, Temperature unit, Valley/ Peak-picker with decay function, Laser operation
Emissivity 0.100 to 1,000 in 0,001-steps
Response time from 30 ms to 10 s (0,03, 0,1, 0,3, 1, 3, 10 s); Option: 5 ms available
Temperature unit °C, °F or K
Analog output - Hardware
Analog output - Functions
linear 0-20mA or 4-20mA, Zoom function for temperature span > 50 °C.
Actual value, Maximum value or Minimum value
Serial interface RS232-Interface, bi directional, 9,600 to 57,600 bps, for programming and data transfer
Thermal switch Monitors the instrument temperature
Power requirements 10.5-32 VDC or 10.8 VAC to 26.4 VAC, 48-400 Hz
Approx. 2.5 W
Permissible operating temperature -25°C ... 60°C
With protective and cooling housing WK15 up to 300 °C
Storage temperature -40°C ... 85°C
Protective class, Weight IP68 (IEC), NEMA 6; 1.3 kg
Housing Stainless steel


Wintronics is the exclusive US agent for HEITRONICS Infrared Radiation Pyrometers

 

 

 

 

Wintronics Calibration provides the highest level of measurement service available today to ensure the exacting quality you demand from your measurement equipment.

 

 

 


Wintronics' Field Engineer Services manages all your engineers' calibration needs - from equipment scheduling and tracking to shipping and calibration - and provides quality reports and certificates of calibration.