CT18
CT18LL
CT18 Brochure
CT18.03 Series Datasheet CT18.03-LL Fiber Optic Series Datasheet
CT18.04 Series Datasheet CT18.04-LL Fiber Optic Series Datasheet
Depends on model, minimum and maximum measuring temperature. (See table above)
Depends on measured temperature and response time, typical value 0.1°C (at 100 ms, 350°C, ε = 1)
+/- 0.5°C plus 0.5% of the difference between target and sensor head temperature
Better than 0.01% of the absolute measured value per month
From 0.4 mm diameter, depends on lens
0.050 to 1.000 in 0.001-steps
From 1 ms to 10 s (0.001, 0.003, 0.01, 0.03, 0.1, 0.3, 1, 2, 10 s)
Linear 0-20 mA or 4-20 mA or 0-10 V scalable temperature span => 200°C
Actual value, max-value, or min-value
RS232-interface, bi-directional, 9.6 kbps to 230 kbps and
RS485 interface, Half-Duplex or Full-Duplex, 9.6 kbps to 230 kbps for programing and data transfer
-20°C to 70°C (-25 to +200 °C for fiber head)
With protective and cooling housing WK15 up to 300°C
-40°C to 85°C
Wintronics Inc., exclusive US agent for HEITRONICS Infrared Radiation Pyrometers
NIST
HEITRONICS
PTB (National Metrology Institute of Germany)
Wintronics, Inc. is an ISO/IEC 17025:2017 Accredited Laboratory that performs Calibration traceable to International Standards.