Field Engineer Calibration:

The Wintronics Field Service System manages all your engineers' calibration needs - from equipment scheduling and tracking to shipping and calibration - and provides quality reports and certificates of calibration.
|

KT-15 Series IIP Infrared Thermometer
Click here to download complete KT15-Series IIP technical specifications, including temperature ranges and resolution tables.
Selection Guide: Infrared Radiation Pyrometer KT15-Series IIP
Model |
Spectral Response/um |
tmin..tmax/°C |
Typical Application |
KT15.01-IIP |
2.00...2.70 |
250...2500 |
metals, metal oxides, ceramics and glass volume |
KT15.02-IIP |
2.00...4.50 |
140...2500 |
metals, metal oxides, glass volume (low temperatures) |
KT15.23-IIP |
6.80±0.15 |
0...400 |
thin film plastics, e.g. PE, PP, PS |
KT15.24-IIP |
7.93±0.15 |
0...400 |
tin film plastics, e.g. PET, Acrylic, PC, PU, FEP, acetate |
KT15.25-IIP |
8.05±0.15 |
0...400 |
thin film plastics, e.g. PTFE, PVC, polyimide, acrylic, PC, PU, FEP, acetate |
KT15.41-IIP |
3.90±0.10 |
250...2500 |
glass volume, measurements through hot gases and flame |
KT15.42-IIP |
4.90...5.50 |
100...2500 |
glass and quartz surface and through hot gases and flame |
KT15.43-IIP |
7.50...8.20 |
0...1400 |
glass and quartz, surface thin ceramics |
KT15.61-IIP |
4.26±0.13 |
400...2500 |
hot gases and flame (CO2-band) |
KT15.62-IIP |
4.50±0.10 |
400...2500 |
hot gases and flame (CO2-and CO-band) |
KT15.63-IIP |
4.66±0.10 |
400...2500 |
hot gases and flame (CO-band) |
KT15.69-IIP |
x...y |
0...1700 |
hot gases in incinerators and fossil fuel fired utility boilers |
KT15.81-IIP
KT15.82-IIP
KT15.83-IIP |
8...10
8...14
8...20 |
0...1000
-30...1000
0…500 |
paper, textiles, rubber, wood, ceramics, thicker plastics (>1 mm)enamel or coated surfaces, asphalt, building materialselectronic components, food, liquids and meteorological, biological, agricultural studies |
KT15.85-IIP |
9.6-11.5 |
-100...200 |
Meteorology, long distance water, ice & cloud temperature measurement |
KT15.99-IIP |
x…y |
-50…3000 |
Various custom spectral bands for customer specific applications |
KT15-Series IIP General Specifications
Temperature range |
Depends on model, minimum and maximum measuring temperature. (See table above)
|
Temperature resolution (NET)* |
Depends on model, detector size, measuring temperature and response time typical value ±0.06°C)
|
Calibration accuracy |
±0.5°C plus 0.7% of the difference between target temperature and housing temperature
|
Long-term stability |
Better than 0.01% of the absolute measured temperature in Kelvin/month
|
Lenses* |
Several far-focus and close-focus lenses are available. See literature Field of view diagrams |
Field of view* |
Depends on model, detector and lens type. For close focus lenses: from 0.7 mm diameter at <20 mm distance; for far-focus lenses, typically 20 to 30 mm diameter at distances >1000 mm
|
Aiming accessories* |
Optional integrated laser pointer or focus laser. Several external optical and mechanical options are available.
See literature Options and Accessories |
Emissivity setting |
Manually adjustable from 0.10 to 1.00, or programmable in 0.001 increments
|
Response time |
Selectable: 50 ms, 100 ms, 300 ms, 1 s, 3 s or 10 s for T90.
Optionally 5mS to 600S
|
Analog output |
4 scalable output signals: linear voltage or current, 0-1(10)V or 0(4)-20 mA. Integrated peak/valley picking w/ programmable decay. Output can be temperature or relative radiance
|
Serial Interface (Option) |
RS-232 interface, bi-directional, baud rate 9600 to 115,000 bps, for free programming and data transfer |
Power requirements |
24 VAC ±10%, 48 to 400 Hz, or 22 to 30 VDC, max. 130 mA with 24 VDC. Optional 9-36VDC, 12-24VAC |
Permissible ambient temperature |
-20 to 60°C (-4 to 140°F) or up to 315°C (600°F) with WK200L cooling jacket
|
Storage temperature |
20°C to 70°C (-4 to 158°F)
|
Housing protection/weight |
IP 65 (DIN 4005) NEMA 4 equivalent, approx. 1.3 kg) |
Wintronics Inc., exclusive US agent for Heitronics Infrared Radiation Pyrometers
|
|